怎樣差異恒溫衰老房與恒溫進行試驗箱?在長時間與使用者進行聯係中,常常進行遭遇使用(yòng)者針對恒溫進行試驗箱與恒溫衰老房(燒機房)等參照近乎實際作用分界糊裏糊塗的設備(bèi)分不清(qīng)的狀況,在此做了其中(zhōng)一個較係統的(de)作出說明。期望對大家掌握恒溫衰老房和恒溫進行(háng)試(shì)驗(yàn)箱有提供(gòng)幫助啊。
How to distinguish high temperature aging room from high temperature test box? In the long-term contact with users, users often encounter ambiguous boundary between high temperature test box and high temperature aging room (burning machine room) and other indicators. Here, a more systematic explanation is made. Hope to help you understand the high temperature aging room and high temperature test box. 恒溫進行試驗箱可用於進行試驗工藝品做恒(héng)溫能夠適應性進行(háng)試驗及恒溫剪應力進行試驗,也可當索(suǒ)然無味箱用,但因為參照(zhào)差別大,索然無(wú)味箱無法代替恒(héng)溫進行(háng)試驗箱(xiāng)。
The high temperature test box can be used for high temperature adaptability test and high temperature stress test of test samples, and also for drying box. However, because of the great difference of indexes, the drying box can not replace the high temperature test box.
恒溫衰老(lǎo)房通常可用於產品出廠前的(de)成熟受到,挑除前一階段無效鑄件。更加多用於農電工價(jià)電子產品。
The high temperature aging room is generally used for the veteran screening of products before they leave the factory, and for the removal of pre-failure workpieces. It is mostly used in electrical and electronic products.
一、與(yǔ)之對(duì)應(yīng)的產品規範標準
1. Corresponding Product Standards
恒溫進行(háng)試驗箱現在落實規範標準GB/T11158-2008《恒溫進行試驗箱技術具體條件》
Existing Implementation Standard GB/T11158-2008 "Technical Conditions for High Temperature Test Box"
恒溫衰老房並未(wèi)特地的產品規範標準,通常參考GB/T11158-2008《恒溫進行試驗箱技術具體(tǐ)條件》落實(shí)
There is no special product standard for high temperature aging room, which is generally implemented with reference to GB/T 1158-2008 "Technical Conditions for High Temperature Test Box"
二、參照的異議
Differences and Similarities of Indicators
恒(héng)溫進行試驗箱
High temperature test box
熱度盛衰興廢度:≤1℃
Temperature fluctuation: <1 Temperature
熱度角度:≤2℃
Temperature gradient: <2 Temperature
熱度表現誤差:±2℃
Temperature deviation: +2 C
恒溫衰老房
High temperature aging room
因為恒溫衰老房並未產品規範(fàn)標(biāo)準,且采用(yòng)熱度不高,通常參考恒溫進行試驗箱參照:
Because there is no product standard for high temperature aging room and the use temperature is not high, generally refer to the index of high temperature test box:
熱(rè)度盛衰興廢度(dù):≤1℃
Temperature fluctuation: <1 Temperature
熱度平均度:≤2℃(按GB/T11158-98的參照(zhào))
Temperature uniformity: <2 C (according to GB/T11158-98)
熱度表(biǎo)現誤差:±2℃
Temperature deviation: +2 C
換(huàn)耐受力(lì):持續可調
Air exchange capacity: adjustable
三、整體結構
Three, structure
因為恒溫進行(háng)試驗箱溫場具體要求較(jiào)高(gāo),通常使用層流設計。
The laminar flow design is generally adopted because of the high temperature test box temperature field requirements.
恒溫衰老房通常要全麵考慮試品架或試品(pǐn)車的狀況。
High temperature aging room should generally consider the situation of test rack or test car.